Review Article
DOI: 10.1016/j.jmrt.2018.08.003
Open Access
Mechanical properties, corrosion resistance and biocompatibilities of degradable Mg-RE alloys: A review
Visits
432
Dexue Liua,b,
Corresponding author
dexeliu@hotmail.com

Corresponding author at: State Key Laboratory of Advanced Processing and Recycling of Non-ferrous Metals School of Material Science and Engineering, Lanzhou University of Technology, Langongping Road 287, Lanzhou City, Gansu Province 730050, China.
, Donglin Yanga, Xinling Lia, Shiwen Hua
a School of Material Science and Engineering, Lanzhou University of Technology, Lanzhou 730050, China
b State Key Laboratory of Advanced Processing and Recycling of Non-ferrous Metals, Lanzhou 730050, China
This item has received
432
Visits

Under a Creative Commons license
Received 04 April 2018, Accepted 25 August 2018
Article information
Abstract
Full Text
Bibliography
Download PDF
Statistics
Figures (8)
Show moreShow less
Tables (1)
Table 1. Cell viability of several cell lines cultured in Mg-RE alloys and its alloys extracts.
Abstract

Magnesium-rare earth (Mg-RE) alloys, as potential biodegradable materials, are attracting considerable attention owing to their good mechanical properties and biodegradation behavior. Indeed, some RE elements within certain amounts of addition have proved to exhibit biocompatibility. Two main strategies have been applied to develop biomedical Mg-RE alloys. One strategy depends on the industrial fabrication system for these alloys while the other is to redesign them from the view point of biomaterials. In this work, the advances in new Mg-RE alloys design are reviewed. It is also summarized the mechanical properties, biocorrosion resistance and biocompatibilities of currently reported Mg-RE alloys. Then, several Mg-RE alloys that may be used in future medical applications are proposed. Finally, a strategy for ongoing development of biomedical Mg-RE alloys is suggested.

Keywords:
Mg-RE alloys
Mechanical property
Biodegradation
Biocompatibility
Full Text
1Introduction

Magnesium based alloys and their use in medical implants have been the focus of significant recent research. Compared to conventional metallic implant materials, such as titanium alloys, stainless steels, and cobalt alloys, biomedical magnesium alloys have better degradability and absorbability [1]. Additionally, magnesium alloy implants have excellent biocompatibility and can enhance bone formation [2]. However, insufficient strength and a significant degradation rate are major obstacles to the use of magnesium in implant applications, as these factors can decrease the structural integrity of the implant, allowing the formation of gas cavities before the tissue is completely recovered [3]. Therefore, pure magnesium must be supplemented with other elements to improve the mechanical property [4] and corrosion resistance of these materials [5,6]. Rare earth (RE) elements are promising candidates as supplemental elements for magnesium alloys used in biodegradable devices for biomedical applications [7,8]. Many studies have shown that rare earth-containing magnesium alloys can be used in biomedical applications due to their excellent biocompatibility, appropriate mechanical properties, and good corrosion resistance [9]. However, some rare earth elements, such as lanthanum, cerium and praseodymium, must be used cautiously in biomedical applications due to their potential toxicity [8,10].

2Alloy strategy for biomedical magnesium alloys

An important step of current research is to optimize alloy design and processing to obtain magnesium-based biomaterials with excellent properties.

2.1Mg-RE binary alloys

Several rare earth elements have been studied in binary magnesium alloys, including lanthanum (La), cerium (Ce), neodymium (Nd), gadolinium (Gd), dysprosium (Dy), yttrium (Y) and scandium (Sc). The addition of trace La and Ce elements effectively improved corrosion resistance and the plasticity of the resulting alloys, but in vitro tests indicated that materials prepared with La and Ce resulted in higher toxicity than ones made with other elements [8,10]. In vivo tests indicated that these elements localized in a small area around the implant [8,11]. According to Willbold et al. [8], no gas cavities appeared in the bodies of rabbits with Mg-La and Mg-Ce alloys implants, and the operation sites were recovered well after four weeks. As heavy RE elements, Gd and Dy have higher solid solubility, 23.49% and 25.3% respectively, than that of Nd, indicating these elements may be more suitable for use in alloys [10]. Hort et al. [12] found that the property profile of Mg-Gd alloys was much closer to the values of bone and the elongation of alloys was better than that of other metallic implant materials, like stainless steels, titanium alloys, and cobalt-chromium alloys. The high solubility of Gd in Mg allowed extremely wide ranges of the mechanical properties to meet different application requirements [12]. However, Myrissa et al. [13] studied the corrosion behavior of Mg-10Gd pins implanted in rats, and found that Gd concentration increased with the implantation time in some animal organs during a 36-week experiment, especially in the spleen, liver, lung and kidney. About Mg-Dy binary alloy, Yang et al. [14,15] found that as-cast Mg-10Dy alloy exhibited appropriate mechanical (ultimate tensile strength (UTS) up to 131MPa, yield strength (YS) up to 83MPa) and corrosion properties (0.8mm/year in 0.9% NaCl) for use in biomedical implants. The in vitro result showed that Mg-10Dy alloys prepared by solutionizing heat treatment (T4) and artificial aging heat treatment (T6) exhibited excellent corrosion resistance (1.34 and 0.97mm/year, respectively) in cell culture medium due to the formation of a relatively Dy-enriched film that was more resistant to corrosion. In binary magnesium alloys, Y addition has shown great hope for degradable implant applications [16]. Wang et al. [17] found that Y ion implantation of pure magnesium improved the resulting corrosion resistance. Peng et al. [18] investigated an effective zone solidification method to prepare high purity Mg-Y biomaterials, and the result showed that the corrosion resistance of Mg-8Y alloy was greatly improved compared to that of as-cast alloys lacking Y addition (2.17mm/year vs 7.11mm/year). As an alloy additive, Sc has been shown to exhibit a grain refining effect that is better than that of other rare earth elements like Y and Ce [19]. Brar et al. [20] studied as-cast Mg-3x alloys (x=Gd, Y, Sc). Compared with Y and Gd, there was better grain refinement by Sc (up to 119μm) and the Mg-3Sc alloy showed the lowest corrosion rate (1.01ml/cm2/day) in Hanks’ solution.

2.2Classic Mg-RE alloys

Mg-Zn based, Mg-Ca based, Mg-Al based, Mg-Li based and WE systems have been considered for use in materials to be implanted in the human body. Screws, pins and stents made of commercial WE43 (W represents Y element, E represents RE elements, 4wt.% Y, 3wt.% RE) are currently being tested in clinical trials [21].

2.2.1Mg-Zn-RE ternary alloys/Mg-Zn-X-RE quaternary alloys

Zn exists in all human tissues and is one of the most abundant nutritionally essential elements in the human body [21]. Zn is commonly used as an alloying element in magnesium alloys to improve the mechanical properties of the alloy, providing a maximum solubility of 6.2wt.%. The addition of rare earth elements to Mg-Zn alloys is considered an effective way to weaken and modify the basal texture of magnesium alloys to obtain excellent properties [22,23]. Thus, a variety of Mg-Zn-RE alloys, such as Mg-Zn-Gd [22] and Mg-Zn-Y [23] alloys, were developed with high strength, ductility balance, and excellent corrosion resistance. Bian et al. [24] developed three magnesium alloys with both Gd and Zn added together at low levels. Of these, Mg-1.8Zn-0.2Gd alloy exhibited a relatively low level of corrosion rate (<0.28mm/year) and no cytotoxicity in L929, MG63, and VSMC cells in vitro (cell viability of higher than 80% for all) and the implant maintained its structural integrity for two months, and only some residual parts were detected after 6 months in vivo, demonstrating the potential use of Mg-Zn-Gd alloys in implant materials. Miao et al. [25] also reported that Mg-2.4Zn-0.8Gd alloy was tolerant of solution treatment. Additionally, extrusion resulted in high mechanical properties (yield strength up to 284MPa, ultimate tensile strength up to 338MPa, and elongation up to 24%) and good corrosion resistance in Hanks’ solution (corrosion rate of 0.16mm/year measured by hydrogen evolution or 0.21mm/year measured by the mass loss) due to a homogeneous microstructure with uniformed grain size and dispersed secondary phases [25]. In studies of Mg-Zn-Y alloys, Kawamura et al. [26] found long period stacking ordered (LPSO) phase in ternary Mg-Zn-Y alloys, with high strength and usable ductility. Zhao et al. [27] further studied the WZ21 (2wt.% Y, 1wt.% Zn) alloy after it was micro alloyed with zirconium and warm extrusion. The result showed that LPSO exhibited high mechanical strength and low degradation rate, suitable characteristics for use as a candidate material for biodegradable orthopedic implants.

Zhang et al. [28] found that increased Y content decreased the grain size of as-cast and as-extruded Mg-Zn-Y alloys (80–200μm) but increased the tensile properties (UTS 260–266MPa, YS 198–213MPa) and elongation (23%-25%). Immersion tests in Hanks’ solution indicated that alloys with single I-phase (Mg3Zn6Y) as the second phase showed the lowest corrosion rate (<0.1mg/cm2/h). Kraus et al. [29] compared two biodegradable magnesium alloys, ZX50 and WZ21, implanted in the bones of rats. WZ21 persisted a longer time (more than 4 weeks) in the bone, and also generated enhanced bone neoformation around the implant, suggesting the good biocompatibility of WZ21. Hänzi et al. [30] studied the degradation performance and biological response of biodegradable WZ21 and ZW21 (2wt.% Zn, 1wt.% Y) alloys both in vivo and in simulated body fluid (SBF). Immersion testing showed that the Mg-Zn-Y alloys exhibited fairly homogeneous degradation behavior and corrosion rates were similar to that of WE43 (about 3.5ml/cm2). An assay of cytocompatibility showed that an eluted pH value of ZW21 of 9.0, compared to 8.5 and 8.25 for WZ21 and WE43, respectively. After 27 days of in vivo testing, some hydrogen bubbles were observed, with significantly reduced gas after 91 days. These results suggested that WZ21 exhibits good biodegradable due to the addition of Y. Gao et al. [31] investigated a new Mg-5.6Zn-0.55Zr-0.9Y alloy by adding Y into ZK60 and found that the addition of the alloying element Y improved the corrosion resistance of Mg-Zn alloys (1.7% vs 3.1%, mass loss after 242h in SBF).

ZEK100 (1wt.% Zn, 0–0.5wt.% Zr, 0–0.5wt.% RE) was found to be a potential biodegradable implant material for osteosynthesis due to its good mechanical properties [32,33]. Waizy et al. [32] investigated the corrosion of ZEK100 in Hank's Balanced Salt Solution (HBSS). The bending strength of the plates was lowest after 4 weeks impression (0.05Nm), compared to 2 weeks (0.08Nm) and 6 weeks (0.1Nm). The highest amounts of O, P, and Ca were detected on the surface, indicating the precipitation of calcium phosphates on the surface of ZEK100 plates. Weizbauer et al. [34] compared the corrosion behavior of ZEK100 and MgCa0.8 in HBSS. The corrosion rates of ZEK100 were significantly lower than that of MgCa0.8 after 4 days (0.63mm/year vs 2.89mm/year). The bending strength of ZEK100 (0.306Nm) was stronger than that of MgCa0.8 (0.272Nm) in the initial state. After 4 days, the bending strength of both alloys decreased by 7%. Huehnerschulte et al. [35], Dziuba et al. [33] and Reifenrath et al. [36] investigated the degradation behavior of ZEK100 in vivo. Huehnerschulte found that the degrading ZEK100 caused adverse host reactions by inducing the unfavorable osteoclastogenic resorption of bone and a rushed reactive formation of new bone periosteal after 6 months. Dziuba conducted a 9–12 month experiment, and found that ZEK100 had pathological effects on the host tissue following complete degradation. Reifenrath performed studies 4–6 weeks after implantation and observed a weight reduction of 7.5% and pull-out forces that were significantly decreased to 44.4% compared to the original value. These different in vivo tests suggest that ZEK100 is suitable for implant applications due to its very high initial stability and a positive biodegradation rate [33,36].

Nd element was also shown to improve mechanical properties and corrosion resistance with acceptable toxicity [37]. JDBM alloy (Mg-(2.0–4.0) Nd-(0.1–0.5) Zn-(0.3–0.6) Zr, wt.%) has high mechanical property (UTS and YS of 300MPa, elongation increased 15%) and excellent corrosion resistance (0.0125mm/year in SBF) [37,38]. Niu et al. [39] found the JDBM alloy could be absorbed in vivo after 18 months (with an average corrosion rate of 0.122mm/year).

Some references suggested that magnesium alloy containing 1wt.% Zn and 1wt.% Mn was easily absorbed in vivo as a bone implant [40]. Stulikova et al. [41] prepared and investigated Mg-4Y/4Ce-1Zn-1Mn alloys, in the T5 condition, and found that the equilibrium Mg12Ce and Mg24Y5 phase enhanced the mechanical properties of the alloys (UTS of about 180MPa and YS of about 130MPa). He et al. [42] studied the corrosion behavior of extruded Mg-Zn-Mn-Y alloy in Hank's solution and found that a protective Y2O3 film formed on the surface of the alloy, improving the bio-corrosion properties (pH value 7–8).

Ca is the most important element of the human skeleton and is the preferred choice for alloying [43]. Zhang et al. [44] and Tong et al. [45] investigated the effect of La/Ce on the microstructure, mechanical properties, and corrosion performances of an as-cast/solid solution Mg-Zn-Ca alloy and an extruded Mg-Zn-Ca alloy. The results showed Ce/La effectively refined different states of Mg-Zn-Ca alloys and restrained grain growth (up to 2–5μm). The addition of 0.5% Ce/La improved the mechanical property of the extruded Mg-Zn-Ca alloy (UTS of 311MPa, YS of 270MPa), while further addition reversed the effect. The corrosion resistance was improved with the increase of Ce/La from 0.5% to 1% (30 to 20mm/year).

2.2.2Mg-Ca-RE alloys/Mg-Ca-Zr-RE alloys

Ca is a major component in human bone and also beneficial to grain refinement of magnesium alloys. A recent study reported that the addition of Ca increased magnesium tensile property and corrosion resistance [46], suggesting better properties might be obtained by adding rare earth elements to the Mg-Ca alloy. Li et al. [46] developed a Mg-1Ca-1Y alloy as a potential biomaterial and reported increased elongation (up to 15.9%) and decreased corrosion resistance (up to 100μg/mm2/day) for the Mg-Ca alloy with Y addition. You et al. [47] investigated the effect of Y addition on a Mg-Ca-Y-Zr alloy, and found that the strength was largely improved with increased addition of Y (UTS of 145–191MPa, YS of 66–120MPa, and elongation of 5.4%-8.3%).

2.2.3Mg-Al-RE alloys

The addition of Al element can effectively promote grain refinement and improve the corrosion resistance of magnesium alloys through solution and precipitation strengthening [48,49]. Yokobayashi et al. [50] prepared a new ternary Mg-Al-Gd alloy by high frequency induction melting and observed two types of needle-like precipitation, including Mg5Gd and LPSO phases after heat treatment. Lu et al. [51] found that heat-treated (T4 and T6) alloy exhibited better corrosion resistance in 1wt.% NaCl solution (331μm/year vs 226μm/year) due to the presence of these needle-like precipitations on the surface. However, Al can harm neurons and enhance the risk of Alzheimer's disease [52]. El-Rahman et al. [52] concluded that accumulated aluminum in brain and altered amino acid neurotransmitters are important mechanisms of aluminum neurotoxicity. According to Rao et al. [53], the direct effect of aluminum depends on Al binding to different brain cells including astrocytes, neural cells and synaptosomes, but interactions with astrocytes inhibits activity of membrane-bound Na+, K+, and Ca2+ ATPase, leading to cellular alterations, and even death.

2.2.4Mg-Li based alloys

In the past decade, Mg-Li based alloys have been studied for biomedical applications, especially LAE442 alloy (4wt.% Li, 4wt.% Al, 1.3wt.% Ce, 0.37wt.% Nd, 0.5wt.% La) [54–58]. Witte et al. [54] investigated the corrosion resistance of gravity-casted LAE442 and AZ91D (Mg-9Al-1Zn, wt.%) in vivo and in vitro. AZ91D was more resistance to corrosion than LAE442 as determined by in vitro tests (−0.267mm/year vs 5.535mm/year) but LAE442 showed greater corrosion resistance than AZ91D in vivo (1.205×10−4mm/year vs 3.516×10−4mm/year). Witte et al. [55] also investigated the corrosion rates of extruded LAE442 with and without additional MgF2 coating. The extruded LAE442 exhibited low corrosion rates (about 0.4mm/year) in vivo and an acceptable host response, but the corrosion resistance was increased by the MgF2 coating (up to about 0.2mm/year). Krause et al. [57] compared the degradation behavior and mechanical properties of extruded LAE442 and WE43 after 6 months in vivo. The initial strength of WE43 implants (238.05N) was lower than that of LAE442 implants (255.67N), and after 6 months, the loss in strength continued at a constant rate. LAE442 implants degraded slower than WE43 as measured by the degradable volume (22.4% vs 33%), suggesting that LAE442 seems sufficient for an application in weight-bearing bones. Zhou et al. [59] compared the mechanical property, biocorrosion, and in vitro biocompatibility of Mg-Li-Al ternary and Mg-Li-Al-RE quaternary alloys. The addition of rare earth elements enhanced mechanical strength while intermetallic compounds distributed throughout the magnesium matrix decreased corrosion resistance. These Mg-Li-based alloys showed no significant reduction of ECV304 cell viabilities (>75%) except for the Mg-8.5Li-2Al-2RE alloy (<60%). Kramer et al. [60] investigated intramedullary nails made of extruded LAE442 by cyclic fatigue tests in both distilled water and HBSS until implant failure or a limit of 500,000 cycles. The result showed that five implants broke in HBSS within the first 70,000 bending cycles but the implants survived in distilled water after 500,000 cycles, which suggested that LAE442 might not have the required fatigue resistance for load-bearing applications in the human body environment. Recently, Minarik et al. [61] investigated the effect of equal channel angular pressing (ECAP) on in vitro degradation of LAE442 magnesium alloy. The corrosion rates of WE43 processed by twelve cycles of ECAP (12p) were 2.49mg/cm2/day and 0.18mg/cm2/day in 0.1M NaCl and in SBF, respectively. In vitro cytotoxicity tests showed, that the ECAP-treated material allowed 80% viability of ECV304 and VSMC cells after 7 days. It was concluded that ECAP could be used to obtain excellent corrosion resistance of LAE442.

2.2.5WE system alloys

Y added to magnesium alloys for biodegradable applications due to its beneficial effect on alloy corrosion and strength [62]. Due to the addition of Y, WE43 alloy containing 4wt.% Y and 3wt.% RE has excellent corrosion resistance. In the past decade, WE43 materials produced by various processing methods have been tested for biomedical application. Hänzi et al. [63], Chu et al. [64], Liu et al. [65] and Davenport et al. [66] investigated the corrosion performance of heat-treated WE43 in vitro. Hänzi found that the performance of annealed and polished WE43 exhibited parabolic behavior in SBF [63]. Degradation decreased over time due to the deposition of corrosion products on the surface while the sample in the thermally oxidized state exhibited more corrosion resistance at the beginning of immersion due to formation of a Y2O3 layer at the surface during annealing at 500□. Chu reported that the corrosion rate decreased after heat treatment due to interactions between finely dispersed precipitates and due to corrosion reactions [64]. Liu concluded that WE43 materials subjected to extrusion and heat treatment transformed from severe diffusional corrosion to uniform corrosion and the corrosion rate decreased from 0.9mm/a to less than 0.4mm/a in PBS solution [65]. The YS, UTS, and elongation increased from 158MPa to 190MPa, from 224MPa to 282MPa, and from 10% to 37%, respectively. Davenport reported that Y-rich regions in the matrix of as-cast WE43 slowed the propagation of corrosion, indicating that the corrosion rate of heat-treated WE43 decreased due to the more homogenous distribution of Y [66]. Liu et al. [67] investigated the properties of WE43 alloy processed by hot extrusion and high-pressure torsion (HPT) and found that the initial grain size decreased to 12μm and the UTS increased from 244MPa to 256MPa after 5 cycles of HPT. Gu et al. [68] investigated the corrosion fatigue behaviors of as-cast AZ91D and extruded WE43 in SBF. The corrosion fatigue limit of the extruded WE43 alloy was obviously higher than that of as-cast AZ91D alloy in SBF (40MPa vs 20MPa) and in air (110MPa vs 50MPa). Moreover, Jamesh et al. [69] investigated the corrosion fatigue of as-cast ZK60 (6% Zn and 0.5% Zr) and WE43 in Ringer's solution and SBF. The corrosion current density (icorr) was 72μA/cm2 for WE43 and 82μA/cm2 for ZK60 in Ringer's solution, but 409μA/cm2 and 419μA/cm2 respectively in SBF suggesting better corrosion resistance of WE43 compared to ZK60.

In addition, many surface modification technologies have been used to further improve the corrosion and biocompatibility of Mg-RE alloys [70]. Jamesh et al. [71] investigated the effect of Zr and N plasma immersion ion implantation on the corrosion behavior of WE43 alloy in SBF and Cell Culture Medium (DMEM). The icorr of treated WE43 was 12-fold lower than that of untreated WE43 in SBF (29.8μA/cm2vs 368μA/cm2) and was 71-fold lower than that of untreated WE43 in DMEM (0.51μA/cm2vs 36.6μA/cm2), suggesting that Zr and N plasma immersion ion implantation could improve corrosion resistance. Jin et al. [72] found significantly increased corrosion resistance and biocompatibility of WE43 magnesium alloy by neodymium self-ion implantation. A 70nm thick Nd2O3 layer covered the surface after implantation, which could act as a protective barrier. Compared to untreated WE43 samples (555μA/cm2 in SBF and 2.55μA/cm2 in the cell culture medium), the treated WE43 samples showed significantly lower icorr values (11.9μA/cm2 in SBF and 0.5μA/cm2 in the cell culture medium). The viability of MC3T3-E1 preosteoblasts in the Nd-implanted WE43 extract was 102%, higher than 98% of untreated WE43 extract. Imwinkelried et al. [73] investigated the effect of a plasma electrolytic coating on WE43. In vitro testing showed high initial gas release rates of non-coated samples (more than 1ml/cm2/day) that decreased to ∼0.3ml/cm2/day after 4–5 days, but the coated samples nearly released no gas during the 4–8 day, and then increased to ∼0.2ml/cm2/day, lower than the non-coated samples. The average weight loss of coated plates in vitro was about 15mg per week compared to 20mg per week for the non-coated plates. In vivo testing showed large gas bubbles around the non-coated plates after one-week implantation that disappeared at the fourth week. No gas pockets were visible around the coated plates after one-week implantation, and tiny gas pockets appeared around the thread holes at the fourth week that remained throughout the lifetime of the animals. These results suggested improved corrosion resistance of coated WE43. Recently, the use of polypyrrole (PPy) coatings was studied as a way to decrease the degradation rate of WE43 alloy [74]. PPy is an electrically-conductive polymer with good biocompatibility and has been used in biomedical applications related to the development of artificial muscles, neural recording stimulation of nerve regeneration, biosensors, and controlled drug release [75]. Ascencio et al. [75] investigated the corrosion behavior of PPy-coated WE43 alloy in a modified simulated body fluid solution (m-SBF). The result showed that the PPy coating effectively decreased the corrosion rate of WE43 alloy due to the formation of a protective layer after 24hours immersion, and the layer could also absorb CO2 to impede charge transfer. Ye et al. [76] investigated a phosphate modified WE43 magnesium alloy and studied its corrosion and biocompatibility in vitro. Immersion test showed that the hydrogen evolution rate of non-coated WE43 (>4ml/cm2/day) was far more than coated WE43 (about 1ml/cm2/day) at the beginning of 50h, and the rate of two groups became more similar (about 1ml/cm2/day) in the next 160h. L-929 cell viability in coated WE43 extract was 96.21%, much higher than the 34.80% viability for the non-coated material. These tests demonstrated that phosphate treatment is an effective way to improve the corrosion resistance and biocompatibility of WE43 alloy for biomedical applications.

Other Mg-Y-Nd-Zr alloys have also been studied. Hakimi et al. [77] investigated a novel Mg alloy (Mg-6% Nd-2% Y-0.5% Zr (EW62)) by rapid solidification (RS) and studied the stress corrosion cracking resistance of the alloy in simulated electrolytes. Immersion tests were divided into short-term (1 day) and long-term (>10 days). In the short-term tests, the corrosion rate of the RS alloy was 0.5mm/year, significantly lower than the 2.1mm/year of conventionally cast (CC) alloy. In the long-term tests, the corrosion rates of the RS alloy and CC alloy were 0.078mm/year and 0.152mm/year, respectively. In the stress corrosion test, the UTS values of RS alloy (from 274MPa in air to 223–262MPa in corrosive environment) were higher than those of the CC alloy (from 181MPa in air to 141–178MPa in corrosive environment). Therefore, EW62 alloy prepared by rapidly solidification has the potential to be used as a biodegradable implant. Aghion et al. [78] compared corrosion performance of EW10X04 (1.16wt.% Nd, 0.48wt.% Y, 0.48wt.% Zr, 0.43wt.% Ca) and EW10 (1.15wt.% Nd, 0.43wt.% Y, 0.46wt.% Zr) in 0.9%NaCl, and found a lower corrosion rate of EW10X04 (about 0.5mm/year) compared to EW10 (more than 1mm/year). However, the stress corrosion resistance of EW10X04 was lower than EW10 as determined by slow strain rate testing.

3Summary of currently investigated Mg-RE alloys3.1Mechanical property

As shown in Figs. 1 and 2, Mg-RE alloys exhibited a large range of UTS (from 76MPa to 354MPa), YS (from 40MPa to 316MPa) and elongation (from 0.7% to 60%). With the increase of rare earth content in Mg-RE binary alloys, UTS and YS increased and elongation decreased. The strengthening effect of Gd was higher than that of Dy. The YS and UTS of most Mg-RE binary alloys were less than 100MPa and 200MPa respectively and the elongation was less than 10%. Overall, Mg-RE binary alloys have lower sufficient strength than that of WE43 alloy for use as implant materials. The elongation of extruded WE43 was more than 35%. Mg-Zn-Y alloys (Mg-1.73Zn-1.54Y, WZ21, ZW21 and ZEK100) exhibited similar yield strength and ultimate tensile strength to that of WE43 with elongation of only about 28%. However, the yield strength and ultimate tensile strength of JDBM and Mg-2.4Zn-0.8Gd alloys exceed those of WE43 reaching 280MPa and 310MPa respectively, with elongation of 15%-25%. The YS and UTS of Mg-4Y/Ce-1Zn-1Mn alloys were about 120MPa and 170MPa, less than those for WE43. The elongation of Mg-4Ce-1Zn-1Mn alloy reached 60%, indicating that addition of Ce element promoted elongation better than the addition of Y. Moreover, hot extrusion [79] and solution treatment also generally improved the strength of Mg-RE alloys, though solution treatment decreased the strength of Mg-Dy binary alloys and Mg-2.4Zn-0.8Gd alloy. Overall, the results suggest that alterations in alloying or processing can be applied to obtain Mg-RE alloys with the appropriate range of mechanical properties for use in biomedical applications.

Fig. 1.
(0.25MB).

Ultimate tensile strength and yield strength of selected Mg-RE alloys.

Fig. 2.
(0.22MB).

Elongation of selected Mg-RE alloys.

3.2Degradation rate

Fig. 3 shows the corrosion rate and hydrogen evolution of Mg-RE alloys in 0.9%/1% NaCl, in vivo and in vitro. The extruded LAE442 and WE43 alloys exhibited the lowest in vivo corrosion rates. Simultaneously, its strength was reduced 40% when the decreased volume of LAE442 reached 15.2%, during the three months in vivo[57]. The loss in strength of WE43 in the three-point bending test was 22.04% after 3 months of implantation [21]. For orthopedic biomaterials, 3–4 months are typically required from fracture callus formation to new bone formation [21], which means that Mg-RE alloys should provide sufficient strength, and durability for bone restoration. Further efforts should focus on improved degradation and corrosion resistance, using methods such as surface treatment [80,81].

Fig. 3.
(0.23MB).

Corrosion rates of selected Mg-RE alloys in 0.9%/1% NaCl, in SBF, in Hank's and in vivo[91,92].

3.3Biocompatibility

Table 1 shows the cell viability of several cell lines cultured in Mg-RE alloy extracts. According to ISO 10993-5:2009 [82], a reduction of cell viability of more than 30% is considered cytotoxic. Compared to pure Mg, the JDBM, Mg-1.8Zn-0.2Gd, WE43 and LAE442 alloys showed good biocompatibility (low cytotoxicity) with L929. The cell viability of Mg-0.69La, Mg-1.27Ce, and Mg-2.13Nd with MC3T3-E1 was lower than that of pure magnesium. Nevertheless, the cell viabilities of Mg-RE alloys with ECV304 and VSMC are roughly equal, and higher than that of Mg-1.8Zn-0.8Gd alloy. In vivo tests in rabbits showed no abnormal reaction after implantation of Mg-La and Mg-Ce alloys for 4 weeks [8]. The JDBM stent covered by brushite coating maintained its original shape after 4-month implantation, with a lower hemolysis rate and anti-vascular restenosis. As bone materials for rabbits, the Mg-1.8Zn-0.2Gd, WE43, and LAE442 implants were well integrated with the surrounding bone within two months, with mineralized areas larger than the magnesium alloy without rare earth elements (for example, AZ91) [55]. The magnesium ions stimulated the increased release of neurotransmitters such as calcitonin gene-related peptide (CGRP) at the periosteum site to further promote osteogenic differentiation of stem cells in the periosteum [83]. Eventually, a large amount of new bone formed at this site.

Table 1.

Cell viability of several cell lines cultured in Mg-RE alloys and its alloys extracts.

Materials  Formation  Cell line  Culture time (day)  Cell viability (%)  Ref. 
Pure Mg  As-cast  L929  65.7  [6] 
  As-cast  MC3T3-E1  87.5  [6] 
  As-cast  ECV304  76.8  [6] 
  As-cast  VSMC  93.6  [6] 
Mg-0.69La  As-cast  MC3T3-E1  78  [8] 
Mg-1.27Ce  As-cast  MC3T3-E1  51  [8] 
Mg-2.13Nd  As-cast  MC3T3-E1  78  [8] 
Mg-1.8Zn-0.2Gd  As-rolled  L929  82  [24] 
  As-rolled  MG63  100  [24] 
  As-rolled  VSMC  90  [24] 
  As-rolled  ECV304  75  [24] 
Mg-1.8Zn-0.8Gd  As-rolled  L929  56  [24] 
  As-rolled  MG63  110  [24] 
  As-rolled  VSMC  93  [24] 
  As-rolled  ECV304  65  [24] 
JDBM  As-extruded  L929  80  [37] 
LAE442  ECAP  L929  100  [61] 
WE43  Phosphating coated  L929  96  [76] 

Although these magnesium implants showed great biocompatibility at the treatment site in animal experiments, it is important to determine the RE content and evaluate RE effects on other organs. In the design of Mg-RE alloys, the toxicity of rare earth alloying elements must first be considered. The toxicity of several rare earth elements were measured in vivo as the amount giving a 50% lethal dose (LD50) [10]. Ce showed the highest cytotoxicity (LD50: 10mg/kg), followed by La (LD50: 150mg/kg), and other rare earth elements (Gd, Nd, Dy, and Y) showed only a slight reduction of viability (LD50: more than 550mg/kg) [10]. In addition to cytoxicity assays, RE pathology trials have been performed on animals. After GdCl3 injection into the abdominal cavity of mice, the ulcer at the wound site developed into a muscular layer in 7 days [84]. In rabbits, GdCl3 affected the contraction of the small intestine [84]. Du et al. [85] found that after treatment with 1000μmol/L GdCl3 in rat dorsal root ganglion neurons for 96h, the cells did not show apoptosis. Additionally, Gd3+ at 100μmol/L in the extracellular solution inhibited the transient outward potassium current (IA). When DyCl3 was injected into rat veins, there was increased activity of liver RNA polymerase II, with no effects on polymerase I. DyCl3 also exhibited antibacterial effect on E. coli at 300μmol/L concentration [86]. Clinically, lanthanum carbonate is used to treat hyperlipidemia in patients with end-stage renal failure. For patients treated for up to 6 years, the concentration of sputum in the blood did not increase during the initial treatment and after treatment, without any liver poisoning [87]. One day after the intravenous injection of CeCl3, there was no Ce3+ in the blood of rats, but after three days, there was jaundice and elevated transglutaminases, with severe hepatotoxicity [88]. The proliferation of osteoclasts was greatly inhibited in the presence of Nd3+, especially at 1×10−7mol/L [85]. In addition, Nd3+ promoted the differentiation and formation of mineralized matrix nodules of osteoclasts at concentrations of 1×10−8mol/L and 1×10−5mol/L, respectively.

Nevertheless these studies have not completely characterized the biocompatibility of all rare earth elements. Rare-earth elements can exist in magnesium alloys in solid solution, as precipitates, and as oxides, so the release forms, degradation, absorption, and metabolism of these elements in animals require more comprehensive characterization.

4Mg-RE alloys most likely to be used in biomedical applications

Biocompatibility is a crucial determinant of whether Mg-RE alloys can be used for biomedical applications. Implanted Mg-RE alloy material should not cause great cytotoxicity or a strong host reaction. According to in vitro cytotoxicity tests listed in Table 1, Mg-1.8Zn-0.2Gd, LAE442, WE43 and JDBM have little effect or no effect on the viability of L929 cells. Fig. 4 shows that L929 cells in the Mg-Zn-Gd extract group and control were generally equal in number and higher in number than the HP-Mg group (green indicates live cells). Mg-Zn based alloys containing trace Gd element have an acceptable cytotoxicity. In terms of hydrogen evolution in vivo, after WE43 and LAE442 materials were implanted for one week, gas bubbles appeared around both alloys and then disappeared gradually after 2–3 weeks, with no detected adverse effects caused by the gas bubbles in vivo[55]. Subcutaneous gas bubbles were observed on postoperative radio-graphs for 4 weeks, as shown in Fig. 5. In addition, no formation of a hydrogen gas pocket was observed during WE43 implantation [89], and WZ21 implantation only caused limited hydrogen gas after 91 days, with no adverse effects observed in vivo[30]. In conclusion, the hydrogen evolution reaction of Mg-RE alloys was inhibited compared to that of pure Mg. The restoration of bone after the implantation of Mg-RE alloys was the most direct evaluation of biocompatibility. Newly formed bone was observed adjacent to the Mg-1.8Zn-0.2Gd implant, and no abnormal effect was observed for bones tissues surrounding the implant [24]. In addition, Witte et al. [54,55] investigated the corrosion behavior of WE43 and LAE442, compared to that of AZ91D. The result showed that the corrosion layer of these alloys contains a lot of Ca and P and was in direct contact with the surrounding bone tissue, which might slow the corrosion process. However, the AZ91D rod was almost destroyed by severe pitting after 18 weeks of implantation, while the LAE442 rod showed more uniform corrosion. Torroni et al. [89] studied the corrosion behavior of WE43 in a cranial bone sheep model. As shown in Fig. 6, newly formed woven bone was observed due to blood clot chambers that formed between the periosteum (red arrows) and bone in regions surrounding the implants (depicted by white arrows in A and B), with virtually no formation of a hydrogen gas pocket. Niu et al. [39] conducted an 18-month implantation of JDBM screws and observed formation of some new bone around the screw, as shown in Fig. 7(a). As shown in Fig. 7(b), after four months, the screw maintained good integrity, with only 16% volume loss, but showed a volume reduction of nearly 90% at 18 months. In addition, Mao et al. [90] conducted a four-month implantation test of a JDBM stent, and as shown in Fig. 8, the stent still maintained its basic shape without fracture of the tube wall, indicating sufficient strength. Based on these studies, magnesium alloys containing RE elements are shown to be suitable as implant materials in bone surgery and vascular stent applications.

Fig. 4.
(0.21MB).

Live/dead staining of L929 cells after culturing in HP-Mg and Mg-Zn-Gd alloy extracts (100% extract, without dilution) for 3 days, normal culture medium as control (green for the live cells).

Fig. 5.
(0.07MB).

Subcutaneous gas bubbles observed on postoperative radiographs for four weeks after implantation during magnesium implant degradation.

Fig. 6.
(0.42MB).

Low- and high-magnification optical micrographs for (a, c) as-cast WE43 and (b, d) WE43-T5.

Fig. 7.
(0.14MB).

(a) Vertical section of a screw. The residual screw is marked by the red dashed line. (b) The original screw and segmented residual screw at 1, 4, and 18 months post operation.

Fig. 8.
(0.14MB).

(a) Image of implantation of JDBM stent in rabbit abdominal artery. (b) SEM image of JDBM stent. (c) 3D reconstruction image of JDBM stent after 4 months of implantation.

5Concluding remarks

With the continued development of biodegradable magnesium alloys, many reports show that addition of RE results in increased corrosion resistance, grain refinement, and solubility of magnesium, enhancing the mechanical and degradable properties of alloys. The WE43 magnesium alloy is currently being clinically tested in the form of screws, pins, and stents and scientists are exploring the potential of other RE-based magnesium alloys. Specific implant materials may be designed for different functions in different parts of the human body. For example, a biodegradable vascular stent should have the ability to undergo plastic deformation, with YS and UTS above 200MPa and 250MPa respectively, and with at least 20% elongation to fracture [93]. The material must be able to function for at least 6 months. A biodegradable screw requires a YS of at least 250MPa, tensile strength of at least 275MPa, elongation rate of at least 10%, and an offset elastic limit of Rp0.2 at least 250MPa in addition [94]. Different RE elements and the amount added can affect the mechanical properties, degradable behavior, and cytotoxicity of alloy materials. As mentioned earlier, Dy, Gd, and Y have high solubility, while Sc has better corrosion resistance and grain refinement capability [19,20]. Some processing techniques can further improve material performance, such as heat-treatment, severe plastic deformation (SPD) [95], and surface modification. Cytotoxicity tests show that Dy, Y, Nd, and Gd are less toxic than La and Ce. However, the effect of RE elements on the body is related to the degradation of metallic implants and the released form of RE elements, therefore systemic in vivo tests are required. Additionally, because in vivo assays are costly and time-consuming, there is a need to develop proper bio-reactors to mimic in vivo conditions for more rapid testing of materials.

Conflicts of interest

The authors declare no conflicts of interest.

Acknowledgement

Dexue Liu was supported by the National Nature Science Foundation of China under Grant Nos. 51365029 and 51664041. This work was also supported in part by International (regional) cooperation and exchange projects-BRICS STI Framework Programmer, the Fundamental Research Funds for the Universities in Gansu Province, the Program for Support Projects of Science and Technology in Gansu Province under Grant No. 1604GKCA038 and Longyuan Youth Innovation and Entrepreneurship Projects.

References
[[1]]
H. Waizy,J.M. Seitz,J. Reifenrath,A. Weizbauer,F.W. Bach,A. Meyer-Lindenberg
Biodegradable magnesium implants for orthopedic applications
J Mater Sci, 48 (2013), pp. 39-50
[[2]]
L. Yang,Y. Huang,F. Feyerabend,R. Willumeit,C. Mendis,K.U. Kainer
Microstructure, mechanical and corrosion properties of Mg-Dy-Gd-Zr alloys for medical applications
Acta Biomater, 9 (2013), pp. 8499-8508 http://dx.doi.org/10.1016/j.actbio.2013.03.017
[[3]]
M.P. Staiger,A.M. Pietak,J. Huadmai,G. Dias
Magnesium and its alloys as orthopedic biomaterials: a review
[[4]]
H. Mirzadeh
Quantification of the strengthening effect of rare earth elements during hot deformation of Mg-Gd-Y-Zr magnesium alloy
J Mater Res Technol, 5 (2016), pp. 1-4
[[5]]
W.C. Kim,J.G. Kim,J.Y. Lee,H.K. Seok
Influence of Ca on the corrosion properties of magnesium for biomaterials
Mater Lett, 62 (2008), pp. 4146-4148
[[6]]
X. Gu,Y. Zheng,Y. Cheng,S. Zhong,T. Xi
In vitro corrosion and biocompatibility of binary magnesium alloys
[[7]]
Y. Liu,Y. Zheng,B. Hayes
Degradable, absorbable or resorbable—what is the best grammatical modifier for an implant that is eventually absorbed by the body?
Sci China Mater, 60 (2017), pp. 377-391
[[8]]
E. Willbold,X. Gu,D. Albert,K. Kalla,K. Bobe,M. Brauneis
Effect of the addition of low rare earth elements (lanthanum, neodymium, cerium) on the biodegradation and biocompatibility of magnesium
[[9]]
L. Yang,Y. Huang,Q. Peng,F. Feyerabend,K.U. Kainer,R. Willumeit
Mechanical and corrosion properties of binary Mg–Dy alloys for medical applications
Mater Sci Eng B, 176 (2011), pp. 1827-1834
[[10]]
F. Feyerabend,J. Fischer,J. Holtz,F. Witte,R. Willumeit,H. Drücker
Evaluation of short-term effects of rare earth and other elements used in magnesium alloys on primary cells and cell lines
Acta Biomater, 6 (2010), pp. 1834-1842 http://dx.doi.org/10.1016/j.actbio.2009.09.024
[[11]]
Y. Chino,M. Kado,M. Mabuchi
Enhancement of tensile ductility and stretch formability of magnesium by addition of 0.2wt%(0.035at%)Ce
Mater Sci Eng A, 494 (2008), pp. 343-349
[[12]]
N. Hort,Y. Huang,D. Fechner,M. Störmer,C. Blawert,F. Witte
Magnesium alloys as implant materials – principles of property design for Mg-RE alloys
Acta Biomater, 6 (2010), pp. 1714-1725 http://dx.doi.org/10.1016/j.actbio.2009.09.010
[[13]]
A. Myrissa,S. Braeuer,E. Martinelli,R. Willumeit-Römer,W. Goessler,A.M. Weinberg
Gadolinium accumulation in organs of Sprague-Dawley® rats after implantation of a biodegradable magnesium-gadolinium alloy
[[14]]
L. Yang,N. Hort,D. Laipple,D. Höche,Y. Huang,K.U. Kainer
Element distribution in the corrosion layer and cytotoxicity of alloy Mg-10Dy during in vitro biodegradation
Acta Biomater, 9 (2013), pp. 8475-8487 http://dx.doi.org/10.1016/j.actbio.2012.10.001
[[15]]
L. Yang,L. Ma,Y. Huang,F. Feyerabend,C. Blawert,D. Höche
Influence of Dy in solid solution on the degradation behavior of binary Mg-Dy alloys in cell culture medium
Mater Sci Eng C, 75 (2017), pp. 1351-1358
[[16]]
M. Liu,P. Schmutz,P.J. Uggowitzer,G. Song,A. Atrens
The influence of yttrium (Y) on the corrosion of Mg-Y binary alloys
Corros Sci, 52 (2010), pp. 3687-3701
[[17]]
X.M. Wang,X.Q. Zeng,G.S. Wu,S.S. Yao
Yttrium ion implantation on the surface properties of magnesium
Appl Surf Sci, 253 (2006), pp. 2437-2442
[[18]]
Q. Peng,Y. Huang,L. Zhou,N. Hort,K.U. Kainer
Preparation and properties of high purity Mg-Y biomaterials
[[19]]
Y. Lee,A. Dahle,D. St John
The role of solute in grain refinement of magnesium
Metall Mater Trans A, 31 (2000), pp. 2895-2906
[[20]]
H.S. Brar,I.S. Berglund,J.B. Allen,M.V. Manuel
The role of surface oxidation on the degradation behavior of biodegradable Mg-RE (Gd, Y. Sc) alloys for resorbable implants
Mater Sci Eng C, 40 (2014), pp. 407-417
[[21]]
N. Li,Y.F. Zheng
Novel magnesium alloys developed for biomedical application: a review
J Mater Sci Technol, 29 (2013), pp. 489-502
[[22]]
G. Yuan,Y. Liu,W. Ding,C. Lu
Effects of extrusion on the microstructure and mechanical properties of Mg-Zn-Gd alloy reinforced with quasicrystalline particles
Mater Sci Eng A, 474 (2007), pp. 348-354
[[23]]
M. Yamasaki,K. Hashimoto,K. Hagihara,Y. Kawamura
Effect of multimodal microstructure evolution on mechanical properties of Mg-Zn-Y extruded alloy
Acta Mater, 59 (2011), pp. 3646-3658
[[24]]
D. Bian,J. Deng,N. Li,X. Chu,Y. Liu,W. Li
In vitro and in vivo studies on biomedical magnesium low-alloying with elements gadolinium and zinc for orthopaedic implant applications
ACS Appl Mater Inter, 10 (2018), pp. 4394-4408
[[25]]
H. Miao,H. Huang,Y. Shi,H. Zhang,J. Pei,G. Yuan
Effects of solution treatment before extrusion on the microstructure, mechanical properties and corrosion of Mg-Zn-Gd alloy in vitro
Corros Sci, 122 (2017), pp. 90-99
[[26]]
Y. Kawamura,M. Yamasaki
Formation and mechanical properties of Mg97Zn1RE2 alloys with long-period stacking ordered structure
Mater Trans, 48 (2007), pp. 2986-2992
[[27]]
X. Zhao,L.L. Shi,J. Xu
Mg-Zn-Y alloys with long-period stacking ordered structure: in vitro assessments of biodegradation behavior
Mater Sci Eng C, 33 (2013), pp. 3627-3637
[[28]]
E. Zhang,W. He,H. Du,K. Yang
Microstructure, mechanical properties and corrosion properties of Mg-Zn-Y alloys with low Zn content
Mater Sci Eng A, 488 (2007), pp. 102-111
[[29]]
T. Kraus,S.F. Fischerauer,A.C. Hänzi,P.J. Uggowitzer,J.F. Löffler,A.M. Weinberg
Magnesium alloys for temporary implants in osteosynthesis: in vivo studies of their degradation and interaction with bone
Acta Biomater, 8 (2012), pp. 1230-1238 http://dx.doi.org/10.1016/j.actbio.2011.11.008
[[30]]
A.C. Hänzi,I. Gerber,M. Schinhammer,J.F. Löffler,P.J. Uggowitzer
On the in vitro and in vivo degradation performance and biological response of new biodegradable Mg-Y-Zn alloys
Acta Biomater, 6 (2010), pp. 1824-1833 http://dx.doi.org/10.1016/j.actbio.2009.10.008
[[31]]
J.C. Gao,S. Wu,L.Y. Qiao,Y. Wang
Corrosion behavior of Mg and Mg-Zn alloys in simulated body fluid
Trans Nonferrous Met Soc, 18 (2008), pp. 588-592
[[32]]
H. Waizy,A. Weizbauer,C. Modrejewski,F. Witte,H. Windhagen,A. Lucas
In vitro corrosion of ZEK100 plates in Hank's balanced salt solution
Biomed Eng Online, 11 (2012), pp. 1-14 http://dx.doi.org/10.1186/1475-925X-11-1
[[33]]
D. Dziuba,A. Meyerlindenberg,J.M. Seitz,H. Waizy,N. Angrisani,J. Reifenrath
Long-term in vivo degradation behaviour and biocompatibility of the magnesium alloy ZEK100 for use as a biodegradable bone implant
Acta Biomater, 9 (2013), pp. 8548-8560 http://dx.doi.org/10.1016/j.actbio.2012.08.028
[[34]]
A. Weizbauer,C. Modrejewski,S. Behrens,H. Klein,P. Helmecke,J.M. Seitz
Comparative in vitro study and biomechanical testing of two different magnesium alloys
J Biomater Appl, 28 (2014), pp. 1264-1273 http://dx.doi.org/10.1177/0885328213506758
[[35]]
A.T. Huehnerschulte,J. Reifenrath,B.V. Rechenberg,D. Dziuba,J.M. Seitz,D. Bormann
In vivo assessment of the host reactions to the biodegradation of the two novel magnesium alloys ZEK100 and AX30 in an animal model
Biomed Eng Online, 11 (2012), pp. 14 http://dx.doi.org/10.1186/1475-925X-11-14
[[36]]
J. Reifenrath,N. Angrisani,N. Erdmann,A. Lucas,H. Waizy,J.M. Seitz
Degrading magnesium screws ZEK100: biomechanical testing, degradation analysis and soft-tissue biocompatibility in a rabbit model
Biomed Mater, 8 (2013), pp. 45012-45021
[[37]]
X. Zhang,G. Yuan,J. Niu,P. Fu,W. Ding
Microstructure, mechanical properties, biocorrosion behavior, and cytotoxicity of as-extruded Mg-Nd-Zn-Zr alloy with different extrusion ratios
J Mech Behav Biomed, 9 (2012), pp. 153-162
[[38]]
X. Zhang,G. Yuan,L. Mao,J. Niu,W. Ding
Biocorrosion properties of as-extruded Mg-Nd-Zn-Zr alloy compared with commercial AZ31 and WE43 alloys
Mater Lett, 66 (2012), pp. 209-211
[[39]]
J. Niu,M. Xiong,X. Guan,J. Zhang,H. Huang,J. Pei
The in vivo degradation and bone-implant interface of Mg-Nd-Zn-Zr alloy screws: 18 months post-operation results
Corros Sci, 113 (2016), pp. 183-187
[[40]]
L. Xu,G. Yu,E. Zhang,F. Pan,K. Yang
In vivo corrosion behavior of Mg-Mn-Zn alloy for bone implant application
J Biomed Mater Res, 83A (2007), pp. 703-711
[[41]]
I. Stulikova,B. Smola
Mechanical properties and phase composition of potential biodegradable Mg-Zn-Mn-base alloys with addition of rare earth elements
Mater Charact, 61 (2010), pp. 952-958
[[42]]
W. He,E. Zhang,K. Yang
Effect of Y on the bio-corrosion behavior of extruded Mg-Zn-Mn alloy in Hank's solution
Mater Sci Eng C, 30 (2009), pp. 167-174
[[43]]
Z. Li,X. Gu,S. Lou,Y. Zheng
The development of binary Mg-Ca alloys for use as biodegradable materials within bone
[[44]]
Q. Zhang,L. Tong,L. Cheng,Z. Jiang,J. Meng,H. Zhang
Effect of Ce/La microalloying on microstructural evolution of Mg-Zn-Ca alloy during solution treatment
J Rare Earth, 33 (2015), pp. 70-76
[[45]]
L.B. Tong,Q.X. Zhang,Z.H. Jiang,J.B. Zhang,J. Meng,L.R. Cheng
Microstructures, mechanical properties and corrosion resistances of extruded Mg-Zn-Ca-xCe/La alloys
J Mech Behav Biomed, 62 (2016), pp. 57-70
[[46]]
Y. Li,P.D. Hodgson,C. Wen
The effects of calcium and yttrium additions on the microstructure, mechanical properties and biocompatibility of biodegradable magnesium alloys
J Mater Sci, 46 (2011), pp. 365-371
[[47]]
S. You,Y. Huang,K.U. Kainer,N. Hort
Influences of yttrium content on microstructure and mechanical properties of as-cast Mg-Ca-Y-Zr alloys
TMS Meeting Exhibition, pp. 91-97
[[48]]
P.T. Ram
Effect of synthetic graphite and activated charcoal addition on the mechanical, microstructure and wear properties of AZ81 Mg alloys
J Mater Res Technol, 5 (2016), pp. 259-267
[[49]]
G. Venkateswarlu,A.K. Singh,J. Davidson,G.R. Tagore
Effect of microstructure and texture on forming limits in friction stir processed AZ31B Mg alloy
J Mater Res Technol, 2 (2013), pp. 135-140
[[50]]
H. Yokobayashi,K. Kishida,H. Inui,M. Yamasaki,Y. Kawamura
Enrichment of Gd and Al atoms in the quadruple close packed planes and their in-plane long-range ordering in the long period stacking-ordered phase in the Mg-Al-Gd system
Acta Mater, 59 (2011), pp. 7287-7299
[[51]]
F. Lu,A. Ma,J. Jiang,Y. Guo,D. Yang,D. Song
Significantly improved corrosion resistance of heat-treated Mg-Al-Gd alloy containing profuse needle-like precipitates within grains
Corros Sci, 94 (2015), pp. 171-178
[[52]]
S.S.A. El-Rahman
Neuropathology of aluminum toxicity in rats (glutamate and GABA impairment)
Pharmacol Res, 47 (2003), pp. 189-194
[[53]]
K.S. Rao
Effect of aluminum (Al) on the brain cells of the rat
Biochem Int, 28 (1992), pp. 51-56
[[54]]
F. Witte,J. Fischer,J. Nellesen,H.-A. Crostack,V. Kaese,A. Pisch
In vitro and in vivo corrosion measurements of magnesium alloys
[[55]]
F. Witte,V. Kaese,H. Haferkamp,E. Switzer,A. Meyer-Lindenberg,C.J. Wirth
In vivo corrosion of four magnesium alloys and the associated bone response
[[56]]
W.-D. Mueller,M.F.L.D. Mele,M.L. Nascimento,M. Zeddies
Degradation of magnesium and its alloys: dependence on the composition of the synthetic biological media
J Biomed Mater Res A, 90A (2009), pp. 487-495
[[57]]
A. Krause,N. Höh,D. Bormann,C. Krause,F.W. Bach,H. Windhagen
Degradation behaviour and mechanical properties of magnesium implants in rabbit tibiae
J Mater Sci, 45 (2009), pp. 624-632
[[58]]
F. Witte,J. Fischer,J. Nellesen,C. Vogt,J. Vogt,T. Donath
In vivo corrosion and corrosion protection of magnesium alloy LAE442
Acta Biomater, 6 (2010), pp. 1792-1799 http://dx.doi.org/10.1016/j.actbio.2009.10.012
[[59]]
W.R. Zhou,Y.F. Zheng,M.A. Leeflang,J. Zhou
Mechanical property, biocorrosion and in vitro biocompatibility evaluations of Mg-Li-(Al)-(RE) alloys for future cardiovascular stent application
[[60]]
M. Krämer,M. Schilling,R. Eifler,B. Hering,J. Reifenrath,S. Besdo
Corrosion behavior, biocompatibility and biomechanical stability of a prototype magnesium-based biodegradable intramedullary nailing system
Mater Sci Eng C, 59 (2016), pp. 129
[[61]]
P. Minárik,E. Jablonská,R. Král
Effect of equal channel angular pressing on in vitro degradation of LAE442 magnesium alloy
Mater Sci Eng C, 73 (2017), pp. 736-742
[[62]]
M. Socjusz-Podosek,L. Lityńska
Effect of yttrium on structure and mechanical properties of Mg alloys
Mater Chem Phys, 80 (2003), pp. 472-475
[[63]]
A.C. Hänzi,P. Gunde,M. Schinhammer,P.J. Uggowitzer
On the biodegradation performance of an Mg-Y-RE alloy with various surface conditions in simulated body fluid
Acta Biomater, 5 (2009), pp. 162-171 http://dx.doi.org/10.1016/j.actbio.2008.07.034
[[64]]
P.W. Chu,E.A. Marquis
Linking the microstructure of a heat-treated WE43 Mg alloy with its corrosion behavior
Corros Sci, 101 (2015), pp. 94-104
[[65]]
D. Liu,Y. Ding,T. Guo,X. Qin,C. Guo,S. Yu
Influence of fine-grain and solid-solution strengthening on mechanical properties and in vitro degradation of WE43 alloy
[[66]]
A.J. Davenport,C. Padovani,B.J. Connolly,N.P.C. Stevens,T.A.W. Beale,A. Groso
Synchrotron X-ray microtomography study of the role of Y in corrosion of magnesium alloy WE43
Electrochem Solid-State Lett, 10 (2007), pp. C5-C8
[[67]]
D.X. Liu,X. Pang,D.L. Li,C.G. Guo,J. Wongsa-Ngam,T.G. Langdon
Microstructural evolution and properties of a hot extruded and HPT-processed resorbable magnesium WE43 alloy
Adv Eng Mater, 19 (2017), pp. 1700723
[[68]]
X.N. Gu,W.R. Zhou,Y.F. Zheng,Y. Cheng,S.C. Wei,S.P. Zhong
Corrosion fatigue behaviors of two biomedical mg alloys – AZ91D and WE43 – in simulated body fluid
[[69]]
M.I. Jamesh,G. Wu,Y. Zhao,D.R. Mckenzie,M.M.M. Bilek,P.K. Chu
Electrochemical corrosion behavior of biodegradable Mg-Y-RE and Mg-Zn-Zr alloys in Ringer's solution and simulated body fluid
Corros Sci, 91 (2015), pp. 160-184
[[70]]
G. Wu,J.M. Ibrahim,P.K. Chu
Surface design of biodegradable magnesium alloys – a review
Surf Coat Technol, 233 (2013), pp. 2-12
[[71]]
M.I. Jamesh,G. Wu,Y. Zhao,W. Jin,D.R. Mckenzie,M.M.M. Bilek
Effects of zirconium and nitrogen plasma immersion ion implantation on the electrochemical corrosion behavior of Mg-Y-RE alloy in simulated body fluid and cell culture medium
Corros Sci, 86 (2014), pp. 239-251
[[72]]
W. Jin,G. Wu,H. Feng,W. Wang,X. Zhang,P.K. Chu
Improvement of corrosion resistance and biocompatibility of rare-earth WE43 magnesium alloy by neodymium self-ion implantation
Corros Sci, 94 (2015), pp. 142-155
[[73]]
T. Imwinkelried,S. Beck,T. Iizuka,B. Schaller
Effect of a plasmaelectrolytic coating on the strength retention of in vivo and in vitro degraded magnesium implants
Acta Biomater, 9 (2013), pp. 8643-8649 http://dx.doi.org/10.1016/j.actbio.2012.08.047
[[74]]
S. Geetha,C.R. Rao,M. Vijayan,D.C. Trivedi
Biosensing and drug delivery by polypyrrole
Anal Chim Acta, 568 (2006), pp. 119-125 http://dx.doi.org/10.1016/j.aca.2005.10.011
[[75]]
M. Ascencio,M. Pekguleryuz,S. Omanovic
Corrosion behaviour of polypyrrole-coated WE43 Mg alloy in a modified simulated body fluid solution
Corros Sci, 133 (2018), pp. 261-275
[[76]]
C.H. Ye,Y.F. Zheng,S.Q. Wang,T.F. Xi,Y.D. Li
In vitro corrosion and biocompatibility study of phytic acid modified WE43 magnesium alloy
Appl Surf Sci, 258 (2012), pp. 3420-3427
[[77]]
O. Hakimi,E. Aghion,J. Goldman
Improved stress corrosion cracking resistance of a novel biodegradable EW62 magnesium alloy by rapid solidification, in simulated electrolytes
Mater Sci Eng C, 51 (2015), pp. 226-232
[[78]]
E. Aghion,G. Levy
The effect of Ca on the in vitro corrosion performance of biodegradable Mg–Nd–Y–Zr alloy
J Mater Sci, 45 (2010), pp. 3096-3101
[[79]]
R. Alizadeh,R. Mahmudi,T.G. Langdon
Superplasticity of a fine-grained Mg–9Gd–4Y–0.4Zr alloy evaluated using shear punch testing
J Mater Res Technol, 3 (2014), pp. 228-232
[[80]]
C.L. Chu,X. Han,J. Bai,F. Xue,P.K. Chu
Surface modification of biomedical magnesium alloy wires by micro-arc oxidation
Trans Nonferrous Met Soc China, 24 (2014), pp. 1058-1064
[[81]]
M.F. He,W.B. Hu,S. Zhao,L. Liu,Y.T. Wu
Novel multilayer Mg–Al intermetallic coating for corrosion protection of magnesium alloy by molten salts treatment
Trans Nonferrous Met Soc China, 22 (2012), pp. s74-s78
[[82]]
ISO 10993-5:2009(E)
Biological Evaluation of Medical Devices-Part 5: Tests for in Vitro Cytotoxicity
International Organization for Standardization, (2009)
[[83]]
Y. Zhang,J. Xu,Y.C. Ruan,M.K. Yu,M. O’Laughlin,H. Wise
Implant-derived magnesium induces local neuronal production of CGRP to improve bone-fracture healing in rats
Nat Med, 22 (2016), pp. 1160 http://dx.doi.org/10.1038/nm.4162
[[84]]
T.J. Haley,K. Raymond,N. Komesu,H.C. Upham
Toxicological and pharmacological effects of gadolinium and samarium chlorides
Br J Pharmacol, 17 (2012), pp. 526-532
[[85]]
Z.Q. Du,P. Yang
Comparative study on apoptosis and membrane potassium current of dorsal root neurons induced by rare earth ions of lanthanum, lanthanum and strontium
J Chin Soc Rare Earth, 29 (2011), pp. 496-503
[[86]]
H.I. Sarkander,W.P. Brade
On the mechanism of lanthanide-induced liver toxicity
Arch Toxicol, 36 (1976), pp. 1-17
[[87]]
A.J. Hutchison,M.E. Barnett,R. Krause
Long-term efficacy and safety profile lanthanum carbonate: results for up to 6 years of treatment
Nephron Clin Pract, 110 (2008), pp. c15-c23 http://dx.doi.org/10.1159/000149239
[[88]]
Y. Nakamura,Y. Tsumura,Y. Tonogai,T. Shibata,Y. Ito
Differences in behavior among the chlorides of seven rare earth elements administered intravenously to rats
Fundam Appl Toxicol, 37 (1997), pp. 106-116
[[89]]
A. Torroni,C. Xiang,L. Witek,E.D. Rodriguez,P.G. Coelho,N. Gupta
Biocompatibility and degradation properties of WE43 Mg alloys with and without heat treatment: in vivo evaluation and comparison in a cranial bone sheep model
J Craniomaxillofac Surg, 45 (2017), pp. 2075-2083 http://dx.doi.org/10.1016/j.jcms.2017.09.016
[[90]]
L. Mao,H. Zhou,L. Chen,J. Niu,L. Zheng,G. Yuan
Enhanced biocompatibility and long-term durability in vivo of Mg-Nd-Zn-Zr alloy for vascular stent application
J Alloy Compd, 720 (2017), pp. 245-253
[[91]]
H.S. Brar,J.P. Ball,I.S. Berglund,J.B. Allen,M.V. Manuel
A study of a biodegradable Mg-3Sc-3Y alloy and the effect of self-passivation on the in vitro degradation
[[92]]
L. Xu,E. Zhang,D. Yin,S. Zeng,K. Yang
In vitro corrosion behaviour of Mg alloys in a phosphate buffered solution for bone implant application
J Mater Sci Mater Med, 19 (2008), pp. 1017-1025 http://dx.doi.org/10.1007/s10856-007-3219-y
[[93]]
Homepage BIOTRONIK AG
Resorbable Magnesium Scaffold (RMS) Indicated for de novo coronary artery lesions [DB/OL]
BIOTRONIK AG, (2015)
Available from: www.magmaris.com/en/product-details [cited 2018 June 18]
[[94]]
H. Windhagen,K. Radtke,A. Weizbauer,J. Diekmann,Y. Noll,U. Kreimeyer
Biodegradable magnesium-based screw clinically equivalent to titanium; screw in hallux valgus surgery: short term results of the first; prospective, randomized, controlled clinical pilot study
Biomed Eng Online, 12 (2013), pp. 62 http://dx.doi.org/10.1186/1475-925X-12-62
[[95]]
D. Peláez,C. Isaza,J.M. Meza,P. Fernández-Morales,W.Z. Misiolek,E. Mendoza
Mechanical and microstructural evolution of Mg AZ31 alloy using ECASD process
J Mater Res Technol, 4 (2015), pp. 392-397
Copyright © 2018. Brazilian Metallurgical, Materials and Mining Association
Journal of Materials Research and Technology

Subscribe to our Newsletter

Article options
Tools
Cookies policy
To improve our services and products, we use cookies (own or third parties authorized) to show advertising related to client preferences through the analyses of navigation customer behavior. Continuing navigation will be considered as acceptance of this use. You can change the settings or obtain more information by clicking here.